HICprevent
This award-winning blog supplements the articles in Hospital Infection Control & Prevention.
CDC and FDA Urge Healthcare Providers to Review Cleaning, Sterilizing Procedures for Reusable Devices
September 11th, 2015
Warning that infection control lapses are endangering patients, the Centers for Disease Control and Prevention and the Food and Drug Administration have issued a joint alert calling for healthcare facilities to review policies and practices in cleaning and processing reusable medical devices.
“Recent infection control lapses due to non-compliance with recommended reprocessing procedures highlight a critical gap in patient safety,” the agencies said in a health advisory issued Friday Sept. 11. “Healthcare facilities (e.g., hospitals, ambulatory surgical centers, clinics, and doctors’ offices) that utilize reusable medical devices are urged to immediately review current reprocessing practices at their facility to ensure they are complying with all steps as directed by the device manufacturers, and have in place appropriate policies and procedures that are consistent with current standards and guidelines.”
The agencies cited recent reports of patients being notified that they may be at increased risk for infection due to lapses in basic cleaning, disinfection, and sterilization of medical devices. These events involved failures to follow manufacturers’ reprocessing instructions for critical (e.g., surgical instruments) and semi-critical items (e.g., endoscopes). The CDC and FDA recommend that healthcare facilities arrange for a healthcare professional with expertise in device reprocessing to immediately assess their reprocessing procedures. This assessment should ensure that reprocessing is done correctly, including allowing enough time for reprocessing personnel to follow all steps recommended by the device manufacturer. In addition, other key recommended actions and interventions include:
Training
• Healthcare facilities should provide training to all personnel who reprocess medical devices.
• Training should be required and provided, upon hire or prior to provision of services at the facility, at least once a year thereafter, and when new devices or protocols are introduced (including changes in the manufacturer’s instructions for use during the device’s life cycle).
• Personnel should be required to demonstrate competency with device reprocessing (i.e., trainer observes correct technique) prior to being allowed to perform reprocessing independently.
• Healthcare facilities should maintain current documentation of trainings and competencies.
• If the healthcare facility hires a contractor for device reprocessing, the facility should verify that the contractor has an appropriate training program and that the training program includes the specific devices the healthcare facility uses.
• Copies of manufacturers’ instructions for operating and reprocessing each type of reusable device should be readily available to staff and inspectors. This file should include instructions for use of chemical disinfectants.
Audits
Healthcare facilities should regularly audit (monitor and document) adherence to cleaning, disinfection, sterilization, and device storage procedures. Audits should be conducted in all areas of the facility where reprocessing occurs. Healthcare facilities should provide feedback from audits to personnel regarding their adherence to cleaning, disinfection, and sterilization procedures. Audits should assess all reprocessing steps, including:
• Performing prompt cleaning after use, prior to disinfection or sterilization procedures • Using disinfectants in accordance with manufacturers’ instructions (e.g., dilution, contact time, storage, shelf-life)
• Monitoring sterilizer performance (e.g., use of chemical and biological indicators, read-outs of sterilizer cycle parameters, appropriate record keeping)
• Monitoring automated endoscope reprocessor performance (e.g., print out of flow rate, time, and temperature, use of chemical indicators for monitoring high-level disinfectant concentration)
Infection Control
• Healthcare facilities should allow adequate time for reprocessing to ensure adherence to all steps recommended by the device manufacturer, including drying, proper storage, and transport of reprocessed devices. Considerations should be made regarding scheduling of procedures and supply of devices to ensure adequate time is allotted for reprocessing.
• Healthcare facilities should have protocols to ensure that healthcare personnel can readily identify devices that have been properly reprocessed and are ready for patient use (e.g., tagging system, storage in a designated area).
• Healthcare facilities should have policies and procedures outlining facility response in the event of a recognized reprocessing error or failure. Healthcare personnel should assess the cause of the error or failure and the exposure event in order to determine the potential risk of infection. The procedure should include how patients who might have been exposed to an improperly reprocessed medical device would be identified, notified, and followed.
• Individuals responsible for infection prevention and reprocessing at the healthcare facility should be consulted whenever new devices will be purchased or introduced to ensure that infection control considerations are included in the purchasing decision as well as subsequent implementation of appropriate reprocessing policies and procedures and to ensure that the recommended reprocessing equipment is available at the healthcare facility.
• Healthcare facilities should maintain documentation of reprocessing activities, including maintenance records for reprocessing equipment (e.g., autoclaves, automated endoscope reprocessors, medical washers and washer-disinfectors, water treatment systems), sterilization records (physical, chemical, and biological indicator results), and records verifying high-level disinfectants were tested and replaced appropriately.
• Healthcare facilities should follow manufacturer recommendations for maintenance and repair of medical devices that are used to perform reprocessing functions as well as medical devices that are reprocessed. If healthcare facilities contract maintenance and repair of these devices to third-party vendors, healthcare facilities should verify that these vendors are approved or certified by the manufacturer to provide those services.